Agilent Technologies E4438C Betriebsanweisung Seite 21

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 34
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 20
User’s and Service Guide Supplement 13
E4438C Option HEC
Repeatable Triggering for Multiple Baseband Sources
54815A Infinium Oscilloscope
1. Turn the power On.
2. Select Channel 1 from the setup menu. This can be accomplished by using the channel
keys above each input BNC connector or by clicking the mouse with the pointer on the
right-hand button to enable the graphical interface.
3. Set Scale to 1.0 V/div, dc Coupling, 50 Input and an Offset of 3.0 VDC.
Refer to Figure 8.
Figure 8 Channel Setup on the Oscilloscope
4. Repeat Step 3 for each channel as follows:
Channel 2 Offset 1.0 Volt, select DC
Channel 3 Offset –1.0 V, select DC
Channel 4 Offset –3.0 V, select DC
Seitenansicht 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 33 34

Kommentare zu diesen Handbüchern

Keine Kommentare