Agilent Technologies Option H48 Multiport Test Set Z5623A Spezifikationen Seite 176

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 191
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 175
38
Page 75
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Automatic Port Extensions – Step 2
After coaxial calibration, connect an open or short to portion of fixture
being measured
Perform APE: algorithm measures each portion of fixture and computes
insertion loss and electrical delay
Values calculated by APE are entered into port extension feature
Now, only fixture mismatch remains as source of error
(dominated by coaxial connector).
Coaxial calibration
reference planes
Ports
extended
Open or short
placed at end of each
transmission line
Page 76
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Measurement Results
Without
Automatic Port
Extension
With Automatic
Port Extension
Delay & loss compensation
(values computed by APE)
DUT
Opens
Stepped Impedance
Adjacent
Structure
Shorts
Balanced
Unbalanced
Balanced
Unbalanced
Balanced
Unbalanced
“Quattro”
demo
board
Seitenansicht 175
1 2 ... 171 172 173 174 175 176 177 178 179 180 181 ... 190 191

Kommentare zu diesen Handbüchern

Keine Kommentare