37
Page 73
安捷倫科技高頻元件量測研討會
Feb.23, 2006
APE = Automatic Port Extensions
First solution to apply both electrical delay
and insertion loss to enhance port extensions
First approach to give reasonable alternative to building
in-fixture calibration standards or de-embedding fixture
1
st
APE accounts for loss
and phase of fixture
transmission lines
Page 74
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Automatic Port Extensions – Step 1
First, perform coaxial calibration at fixture connectors to remove errors
due to VNA and test cables
At this point, only the fixture loss, delay and fixture mismatch remain
as sources of error.
Coaxial calibration
reference planes