Agilent Technologies N9360A Bedienungsanleitung Seite 4

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 10
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 3
4
W-CDMA/HSDPA
Modulated Power Measurement
Range: 0 to +36 dBm; (-30 to +36 dBm)
1
Resolution: 0.1 dB
Accuracy: ±0.7 dB (25 ±5 °C), typical =
±
0.4
±1.0 dB (0 to 50 °C)
Range: 53 to 0.01 dBm; (-60 to -30.01dBm)
1
Resolution: 0.1 dB
Accuracy: ±1.0 dB (25 ±5 °C), typical =
±
0.5
±1.5 dB (0 to 50 °C)
Range: 60 to 53.01 dBm; (-70 to -60.01dBm)
1
Resolution: 0.1 dB
Accuracy: ±1.5 dB (25 ±5 °C), typical =
±
0.8
±2.0 dB (0 to 50 °C)
Frequency Error Measurement
Range: 0 to ±500 Hz
Resolution: 0.1 Hz
Accuracy: ± (10 + reference signal) Hz
Input level: 20 to +36 dBm (30 to +36 dBm)
1
EVM Measurement
Range: 0 to 20%
Resolution: 0.01%
Residual EVM: 3.8%
Input level: 20 to +36 dBm; ( 30 to +36 dBm
)
1
ACLR Measurement
Input level: 5 to +36 dBm
Range: 0 to 40 dB (at ±5 MHz), 0 to 48 dB (at ±10 MHz)
(43 dB (at ±5 MHz) and 53 dB (at ±10 MHz))
1
Resolution: 0.01 dB
SEM Measurement
1
Range: 5 to +36 dBm
Margin: > 7.7 dB
1
Requires N9360AU-100 hardware installed in unit
OBW Measurement
Input level: 5 to +36 dBm
Accuracy: < ±100 kHz
Range: 0.00 to 9.99 MHz
Resolution: 0.01 MHz
Sensitivity/BER
Input level: 20 to +36 dBm
Range:
PN9: 0.00 to 25.00%
PN15: 0.00 to 33.33%
Origin Offset Measurement
Range: 10%
Resolution: 0.01%
Residual EVM: 50%
Input level: 20 to +36 dBm
Seitenansicht 3