Agilent Technologies 8922M Betriebsanweisung Seite 169

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 384
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 168
12-3
CW Commands (Measure Subsystem)
FREQuency[:ABSolute]
FREQuency[:ABSolute]
Description Sets the CW ABSolute FREQuency MEASurement attributes.Queries the CW ABSolute
FREQuency MEASurement result.
GPIB unit is HZ.
Display units are GHZ, MHZ, KHZ, HZ;
default unit is MHZ.
Syntax MEASure:CW:FREQuency[:ABSolute]?
MEASure:CW:FREQuency[:ABSolute][:MM] | [:AVG] | [:MET]
Options Refer to Appendices D, F and G.
FREQuency:ERRor
Description Sets/queries the CW FREQuency ERRor MEASurement attributes.
GPIB unit is HZ.
Display units are GHZ, MHZ, KHZ, HZ;
default unit is MHZ.
Syntax MEASure:CW:FREQuency:ERRor?
MEASure:CW:FREQuency:ERRor[:MM] | [:AVG] | [:MET]
Options Refer to Appendices D, F and G.
FREQuency:SELect
Description Selects/queries the CW FREQ SELected measurement
to display.
Syntax MEASure:CW:FREQuency:SELect?
MEASure:CW:FREQuency:SELect <string>
Options ‘CW FREQ’ | ‘ CWFREQERR’