www.agilent.com/comms/lightwave
Spectrum Analysis
Agilent 83453B High-Resolution Spectrometer (cont.)
Agilent Technologies warrants that the specifications listed below will be met under the following instrument operating conditions:
•
operating temperature +20°C to +30°C
•
wavelength within 1520 to 1620 nm
High Resolution Normal Resolution Wide Resolution
Operating Wavelength Range 1440 to 1640 nm 1440 to 1640 nm 1440 to 1640 nm
with 81600B option 200 Tunable Laser
Frequency Span Range 250 MHz to 5 GHz 250 MHz 125 GHz 400 MHz to 12.5 THz
(2 pm to 40 pm) (2 pm to 1 nm) (4 pm to 50 nm)
Absolute Frequency Accuracy
1,2
±1.8 GHz (±15 pm)
with spans ≤12 GHz (100 pm)
Relative Frequency Accuracy
3
Span ≤5 GHz (≤40 pm) ±125 MHz (±1 pm) ±190 MHz (±1.5 pm) –
Span >5 GHz to 125 GHz (0.04 nm to 1 nm) – ±400 MHz (±3.2 pm) –
Span 1 nm to 50 nm – – ±1.2 GHz (±10 pm)4
Frequency Repeatability
over 5 minutes ±50 MHz (±0.4 pm) ±60 MHz (±0.5 pm) ±300 MHz (±2.4 pm)
Nominal Frequency Resolution
4
Span/Number of Trace Points
Minimum Frequency Resolution
4
1 MHz 15 MHz 20 MHz
Power Accuracy
5,6
over full wavelength range at –15 dBm, ±5 dB
7
±2.75 dB ±6 dB
7
with spans <12 GHz (0.1 nm)
Power Repeatability
8
with spans <12 GHz (0.1 nm), over 5 minutes ±10 dB
7
±0.75 dB ±8 dB
7
Power Scale Fidelity
with spans 100 pm (–10 dBm to –45 dBm) ±1 dB ±1 dB –
Polarization Dependence
7
±1.5 dB ±1.5 dB –
Displayed Average Noise Level
7
RMS detection; 20 pm span; –5 dBm – –65 dBm –
Dynamic Range9, 10 at 1550 nm – ≥50 dB –
Spurious Free Dynamic Range
7
≥40 dB ≥40 dB –
Peak Input Power before Saturation
7
with 0 dB attenuation –5 dBm
Maximum Safe Total Input Power +23 dBm
Optical Attenuation Range 0 to 20 dB in 1 dB steps
Optical Return Loss
7
50 dB
1
Verified at 1525 nm, 1550 nm, and 1615 nm.
2
In cases where the marker resolution (which is span/(trace points –1)) exceeds a frequency/wavelength specification, the marker resolution overrides the specification.
3
Assumes constant temperature over measurement interval.
4
Characteristic
5
Does not include PDL, scale fidelity, or repeatability.
6
With unmodulated linewidths <2 MHz.
7
Characteristic
8
Assumes constant temperature over measurement interval.
9
With a –5 dBm level adjusted for optimum polarization.
10
Applies to any fixed input power level.
General Computer Interfacing
Dimensions (without computer): 280 mm H x 325 mm W x 560 mm D Operating System: Windows XP
Weight: 34 kg (75 lbs) Remote Control Compatibility: LAN Interface
Operating Temperature Range: +20°C to +30°C Data Export: Spreadsheet and Word Processor Compatible (CSV)
Humidity: 15 .. 85% Graphics Export: JPEG, Bitmap
Floppy Disk: 3.5 inch 1.44 MB, MS-DOS
Power Requirements CD-ROM: 40 X Maximum Speed
LAN Interface: Ethernet 10/100 Mbit/s
Voltage and Frequency: 110 to 240 V AC, 50 to 60 Hz Graphical User Interface: TFT Monitor, XGA
Max. Power Consumption: 400 W Keyboard/Mouse: PS/2
Additional Interface: Data Acquisition Cards, GPIB
29