1 Quick Reference
12 Model N6705A User’s Guide
The Agilent N6705A DC Power Analyzer – At a Glance
The Agilent N6705A DC Power Analyzer is a multi-functional power
system that combines the functions of a multiple-output DC voltage
source with the waveform/data capturing capability of an
oscilloscope and data logger.
As a multiple-output DC source, the Agilent N6705A provides up to
four configurable outputs. Available power modules have power
levels of 50 W, 100 W, and 300 W, various voltage and current
combinations, and provide a variety of performance features as
described under “Power Module Capabilities”. Each output also has
arbitrary (Arb) waveform generation capability, which lets you
program up to seven predefined voltage waveforms – or define your
own voltage or current waveform.
As a measurement system, the Agilent N6705A displays the average
output voltage and current on a 4 or 5 digit Meter View. Waveforms
can be displayed using the Scope View, which you can adjust using
vertical and horizontal controls. The Data Logger View measures and
charts average and peak voltage and current measurements over an
extended time period.
Source Features
Color-coded display and
output controls
Correspondence between color-coded information on the display and front
panel connectors and keys.
Programmable voltage
and current
Full programming capability is provided for the entire range of output
voltage and current for all power modules.
Low output noise Output noise is <4.5 mV peak-to-peak for autoranging and precision power
modules, which is comparable to linear supplies.
Fast up/down
programming
1.5 millisecond response time from 10% to 90% of the output rating for
autoranging and precision power modules.
Fast transient response Transient response is less than 100 microseconds for autoranging and
precision power modules.
Autoranging capability Autoranging supplies the maximum rated power over a continuous range of
voltage and current settings for autoranging and precision power modules.
Output On/Off sequencing
A turn-on/turn-off delay capability for each output allows output on/off
sequencing.
Front panel binding posts + and – output and + and – sense terminals are provided for each output.
Sense terminals provide 4-wire voltage measurements.
Output protection Outputs have over-voltage, over-current, and over-temperature protection.
Emergency shut-off An emergency stop button to quickly shut down all outputs.