P-Series Wideband Power Sensor Specifications
The P-Series wideband power sensors are designed for use with the P-Series power meters only.
Sensor model Frequency range Dynamic range Damage level Connector type
N1921A 50 MHz to 18 GHz –35 dBm to +20 dBm ( 500 MHz) +23 dBm (average power); Type N (m)
–30 dBm to +20 dBm (50 MHz to 500 MHz) +30 dBm (< 1 µs duration)
(peak power)
N1922A 50 MHz to 40 GHz –35 dBm to +20 dBm ( 500 MHz) +23 dBm (average power); 2.4mm (m)
–30 dBm to +20 dBm (50 MHz to 500 MHz) +30 dBm (< 1 µs duration)
(peak power)
Maximum SWR
Frequency band N1921A /N1922A
50 MHz to 10 GHz 1.2
10 GHz to 18 GHz 1.26
18 GHz to 26.5GHz 1.3
26.5 GHz to 40 GHz 1.5
Sensor Calibration Uncertainty
9
Definition: Uncertainty resulting from non-linearity in
the sensor detection and correction process. This can be
considered as a combination of traditional linearity, cal
factor and temperature specifications and the uncertainty
associated with the internal calibration process.
Frequency band N1921A N1922A
50 MHz to 500 MHz 4.5% 4.3%
500 MHz to 1 GHz 4.0% 4.2%
1 GHz to 10 GHz 4.0% 4.4%
10 GHz to 18 GHz 5.0% 4.7%
18 GHz to 26.5GHz 5.9%
26.5 GHz to 40 GHz 6.0%
Physical characteristics
Dimensions N1921A 135 mm x 40 mm x 27 mm
N1922A 127 mm x 40 mm x 27 mm
Weights with cable Option 105 0.4 kg
Option 106 0.6 kg
Option 107 1.4 kg
Fixed sensor cable lengths Standard 1.5 m (5-feet)
Option 106 3.0 m (10-feet)
Option 107 10 m (31-feet)
9. Beyond 70% Humidity, an additional 0.6% should be added to these values.