Agilent N1911A/N1912A P-Series PowerMeters and N1921A/N1922AWideband Power SensorsData sheet
10Worked ExampleUncertainty calculations for a power measurement (settled, average power)[Specification values from this document are in bold italic,
11Graphical ExampleA. System contribution to measurement uncertainty versus power level (equates to step 6 result/2)Note: The above graph is valid for
Related Literature ListP-Series Power Meters and Power Sensors,configuration guide, literature number 5989-1252ENP-Series Power Meters and Power Senso
2Specification DefinitionsThere are two types of product specifications:Warranted specifications are specifications which are covered bythe product wa
3P-Series Power Meter and Sensor Key System Specifications and Characteristics2 Maximum sampling rate 100 Msamples/sec, continuous samplingVideo bandw
4P-Series Power Meter SpecificationsMeter uncertaintyInstrumentation linearity ± 0.8 %TimebaseTimebase range 2 ns to 100 msec/divAccuracy ±10 ppm Jitt
5P-Series Wideband Power Sensor SpecificationsThe P-Series wideband power sensors are designed for use with the P-Series power meters only.Sensor mode
61 mW Power ReferenceNote: The 1 mW power reference is provided for calibration of E-series and 8480 series sensors. The P-Series sensors are automati
788.5 mm H x 212.6 mm W x 348.3 mm D (3.5 in x 8.5 in x 13.7 in)Net weight N1911A ≤ 3.5 kg (7.7 lb) approximateN1912A ≤ 3.7 kg (8.1 lb) approximateShi
8Characteristic Peak FlatnessThe peak flatness is the flatness of a peak-to-average ratio measurement for various tone-separations for an equal magnit
9Appendix AUncertainty calculations for a power measurement (settled, average power)[Specification values from this document are in bold italic, valu