Agilent Technologies N1911A Bedienungsanleitung Seite 11

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 12
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 10
11
Graphical Example
A. System contribution to measurement uncertainty versus power level (equates to step 6 result/2)
Note: The above graph is valid for conditions of free-run operation, with a signal within the video bandwidth setting on the system.
Humidity < 70%.
B. Standard uncertainty of mismatch.
SWR r SWR r
1.0 0.00 1.8 0.29
1.05 0.02 1.90 0.31
1.10 0.05 2.00 0.33
1.15 0.07 2.10 0.35
1.20 0.09 2.20 0.38
1.25 0.11 2.30 0.39
1.30 0.13 2.40 0.41
1.35 0.15 2.50 0.43
1.40 0.17 2.60 0.44
1.45 0.18 2.70 0.46
1.5 0.20 2.80 0.47
1.6 0.23 2.90 0.49
1.7 0.26 3.00 0.50
Note: The above graph shows the Standard Uncertainty of Mismatch = rDUT. rSensor /2, rather than the Mismatch Uncertainty
Limits. This term assumes that both the Source and Load have uniform magnitude and uniform phase probability distributions.
C. Combine A & B
Expanded Uncertainty , k = 2, = 2. U
C
= . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ± %
System uncertainty contribution - 1 sigma (%)
1.0%
10.0%
100.0%
-35 -30 -25 -20 -15 -10 -5 0 5 10 15 20
Power (dBm)
N1921A: 500 MHz to 10 GHz
N1922A:18 to 40 GHz
Other bands
Standard uncertainty of mismatch - 1 sigma (%)
0.5
0.45
0.4
0.35
0.3
0.25
0.2
0.15
0.1
0.05
0
0 0.1 0.2 0.3 0.4 0.5
r
Sensor
r
DUT
2
U
C
=
(Value from Graph A) + (Value from Graph B)