FORMat
The FORMat subsystem sets the format for data transferred from the multimeter to
the computer using the
MEASure?, READ?, and FETCh? commands.
Subsystem Syntax FORMat
[:DATA] <type>[,<length>]
[:DATA]
FORMat[:DATA] <type>[,<length>] selects the data format and length.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<type>
discrete ASCii | REAL none
<length>
numeric 32 | 64 none
Comments
ASCII readings are transferred in the form ± 1.234567E ± 123. Each reading is
followed by a comma(,). A line feed (LF) and End-Or-
Identify (EOI) follow the last reading. Each reading is 15 bytes.
REAL reading lengths are 32- and 64-bits. The readings are transferred in the
IEEE 488.2-1987 Definite Length Arbitrary Block format. The readings are
preceded by an Arbitrary Block header which consists of:
# <non-zero digit> <block length>
<non-zero digit> indicates the number of digits representing
<block length>, and <block length> indicates the number of 8-bit data bytes
which follow the header. Examples of the header are shown below:
REAL,32 #240 <40 bytes> 10 readings
REAL,64 #18 <8 bytes> 1 reading
*RST Condition: ASCii
Example Setting the Data Format
FORMAT REAL,64 !Format is set to REAL 64.
CONF:VOLT:DC 58.1
Chapter 5 Command Reference 140