Agilent Technologies E1326B Bedienungsanleitung Seite 104

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 291
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 103
Example 1: Setting the
Trigger Count
(stand-alone
multimeter)
In this example, one DC voltage measurement is taken each time an external
trigger occurs. After 10 external triggers (and measurements), the
multimeter returns to the idle state.
10 !Configure the stand-alone multimeter for DC voltage measurements
20 !on its terminals. Set the multimeter to receive 10 external triggers.
30 !Place the multimeter in the wait-for-trigger state.
40 CONF:VOLT:DC
50 TRIG:SOUR EXT
60 TRIG:COUN 10
70 READ?
Example 2: Setting the
Trigger Count
(scanning multimeter)
In this example, the multimeter scans the channel list five times making a total
of 25 measurements. The multimeter is internally triggered as the trigger
source is not changed from that set by
CONFigure (TRIG:SOUR IMM).
10 !Configure the scanning multimeter for DC voltage measurements on
20 !multiplexer channels 100 through 104. Make 5 scans through the channel list.
30 !Place the multimeter in the wait-for-trigger state.
40 CONF:VOLT:DC (@100:104)
50 TRIG:COUN 5
60 READ?
Querying the Trigger
Count
The trigger count setting is queried with the TRIGger:COUNt? and
TRIGger:COUNt? MIN | MAX commands. See Chapter 5 for additional
information.
Chapter 4 Understanding the HP E1326B/E1411B Multimeter 105
Seitenansicht 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 290 291

Kommentare zu diesen Handbüchern

Keine Kommentare