N6705 A, N6705B
Regulatory
compliance EMC Complies with the European EMC directive for Class A test and
measurement products.
Complies with the Australian standard and carries the C-Tick mark.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Electrostatic discharges greater than 1 kV near the I/O connectors may cause
the unit to reset and require operator intervention.
Safety Complies with the European Low Voltage Directive and carries the CE-marking.
This product also complies with the US and Canadian safety standards for test
and measurement products.
Acoustic noise
declaration This statement is provided to Sound pressure Lp < 70 dB(A),
comply with the requirements At operator position and normal operation, according to EN 27779 (Type Test).
of the German Sound Emission Schalldruckpegel Lp <70 dB(A)
Directive, from 18 January 1991. Am arbeitsplatz und normaler betrieb, nach EN 27779 (Typprüfung).
Output terminal
isolation
Maximum rating No output terminal may be more than 240 VDC from any other terminal or chassis ground.
AC input
Nominal input ratings 100 VAC – 240 VAC; 50/60 Hz/400 Hz
Input range 86 VAC – 264 VAC
Power consumption 1500 VA typical
Fuse Internal fuse (not customer accessible)
Weight
N6705 with 4 installed modules Net: 17.3 kg; 38 lbs.
Single-wide power module Net: 1.23 kg; 2.71 lbs
Double-wide power module Net: 2.18 kg; 4.8 lbs
Agilent N6705A and N6705B DC Power Analyzer Mainframes (Continued)
26