Agilent N6751A/N6752A, N6754A and N6761A/N6762A Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C after a 30-minute warm-up period,
with each module’s sense terminals externally jumpered directly to their respective output terminals (local sensing).
Note: Performance specifications for the N6780 SMU modules can be found at www.agilent.com/find/N6780. See the N6780 Series Source/
Measure Units (SMUs) for the N6700 Modular Power System Data Sheet, literature number 5990-5829EN.
N6751A / N6753A N6754A N6761A /
N6752A N6762A
DC output ratings
Voltage 50 V 20 V 60 V 50 V
Current (derated 1% per °C above 40°C) 5 A / 10 A 50 A 20 A 1.5 A / 3 A
Power 50 W / 100 W 300 W 300 W 50 W / 100 W
Output ripple
and noise (PARD)
(from 20 Hz – 20 MHz) CV peak-to-peak 4.5 mV 5 mV 6 mV 4.5 mV
CV rms 350 µV 1 mV 1 mV 350 µV
Load effect
(Regulation)
(for any output load change, Voltage 2 mV 2 mV 2 mV 0.5 mV
with a maximum load-lead Current (@ 0 - 7 V) 2 mA 12 mA 5 mA 30 µA
drop of 1 V per lead) (@ 0 - 50 V) 2 mA 12 mA 5 mA 65 µA
Source effect
(Regulation) Voltage 1 mV 0.5 mV 1.2 mV 0.5 mV
Current 1 mA 5 mA 2 mA 30 µA
Programming
accuracy
(at 23°C ±5°C Voltage high range 0.06% + 19 mV 0.06% + 10 mV 0.06% + 25 mV 0.016% + 6 mV
after 30 minute Voltage low range (≤ 5.5 V) N/A N/A N/A 0.016% + 1.5 mV
warm-up. Applies Current high range 0.1% + 20 mA 0.10% + 30 mA 0.10% + 8 mA 0.04% + 200 µA
from min. to max. Current low range (≤ 100 mA, @ 0 - 7 V) N/A N/A N/A 0.04% + 15 µA
programming range) (≤ 100 mA, @ 0 - 50 V) N/A N/A N/A 0.04% + 55 µA
Measurement
accuracy
(at 23°C ±5°C) Voltage high range 0.05% + 20 mV 0.05% + 10 mV 0.05% + 25 mV 0.016% + 6 mV
Voltage low range (≤ 5.5 V) N/A N/A N/A 0.016% + 1.5 mV
Current high range 0.1% + 4 mA 0.10% + 30 mA 0.10% + 8 mA 0.04% + 160 µA
Current low range (≤ 100 mA, @ 0 - 7 V)
NOTE 1
N/A N/A N/A 0.03% + 15 µA
NOTE 2
(≤ 100 mA, @ 0 - 50 V) N/A N/A N/A 0.03% + 55 µA
Load transient
recovery time
(time to recover to within the settling band following a load change)
• from 60% to 100% and from 100% to 60% of full load for models N6751A & N6761A
• from 50% to 100% and from 100% to 50% of full load for models N6752A-N6754A & N6762A.
Voltage settling band ± 75 mV
NOTE 2
± 30 mV ± 90 mV
NOTE 3
± 75 mV
Time < 100 µs < 100 µs < 100 µs < 100 µs
1
Applies when measuring 409 6 data points (SENSe:SWEep:POINts = 4096).
2
Settling band is ±125 mV for Model N6752A when relay option 761 is installed.
3
Settling band is ±350 mV for Model N6754A when relay option 760 or 761 is installed.