208 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
16 Save/Recall (Setups, Screens, Data)
• Setup Info — setup information (vertical, horizontal, trigger, acquisition,
math, and display settings) is also saved in a separate file with a
TXT extension.
• Invert Grat — the graticule in the image file has a white background
instead of the black backgound that appears on- screen.
• Palette — lets you choose between Color or Grayscale images.
4 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
To save CSV, ASCII XY, or BIN data files
Data files can be saved to an external USB storage device.
1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select
CSV data (*.csv), ASCII XY data (*.csv), or Binary data (*.bin).
2 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See "To navigate storage locations" on
page 214.
raticule Not Inverted
raticule Inverte
When saving screen images, the oscilloscope uses the last menu visited before pressing
the [Save/Recall] key. This lets you save any relevant information within the softkey menu
area.
To save a screen image showing the Save/Recall Menu at the bottom, press the
[Save/Recall] key twice before saving the image.
You can also save the oscilloscope's display image using a web browser. See "Get
Image" on page 249 for details.