174 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
13 Measurements
Base
The Base of a waveform is the mode (most common value) of the lower
part of the waveform, or if the mode is not well defined, the base is the
same as Minimum. The Y cursor shows the value being measured.
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Rising edge overshοot =
local Maximum D Top
Amplitude
× 100
Falling edge overshοot =
Base D local Minimum
Amplitude
× 100
Overshoot
local Maximum
local Minimum
Base
Top
Overshoot