202 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
15 Waveform Generator
Notice that you can select between coarse and fine adjustments for
frequency, period, and width. Also, pushing the Entry knob is a quick
way to toggle between coarse and fine adjustments.
The Settings softkey opens the Waveform Generator Settings Menu which
lets you make other settings related to the waveform generator.
See:
"To output the waveform generator sync pulse" on page 202
"To specify the waveform generator output load" on page 203
"To use waveform generator logic presets" on page 203
"To restore waveform generator defaults" on page 204
To output the waveform generator sync pulse
1 If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the [Wave Gen] key.
2 In the Waveform Generator Menu, press the Settings softkey.
3 In the Waveform Generator Settings Menu, press the Trig Out softkey and
turn the Entry knob to select Waveform Generator Sync Pulse.
Waveform Type Sync Signal Characteristics
All waveforms except
DC and Noise
The Sync signal is a TTL positive pulse that occurs when the waveform
rises above zero volts (or the DC offset value).
DC N/A
Noise N/A