VAR2
Subordinate linear staircase or linear
pulsed sweep. After primary sweep is
completed, the VAR2 unit output is
incremented.
Maximum number of steps: 128
CONSTANT
A source unit can be set as a constant
voltage or current source depending on
the unit.
PULSE
One of the SMUs can be set as a pulse
source.
Pulse width: 0.5ms to 100ms, 100µs
resolution.
Pulse period:
(5ms to 1s ( pulse width + 4ms),
100µs resolution.
SMU pulse setting accuracy
(supplemental information, at fixed range
measurement except multi- channel
measurement):
Width: 0.5% + 50µs
Period: 0.5% + 100µs
Trigger output delay for pulsed
measurement: 0 - 32.7ms with 100µs
resolution (< pulse width).
Sampling (Time Domain)
Measurement Characteristics
Displays the time sampled voltage/
current data versus time.
Maximum sampling points: 10,001
(linear)
Sampling mode: linear, log, and
thinned-out
Note: The thinned-out mode is similar to
reverse-log sampling. Sampling
measurement continues by thinning out
older data until the sampling completion
condition is satisfied.
Sampling interval range and resolution:
Linear scale (auto mode):
60µs to 480µs range: 20µs resolution
480µs to 1s range: 80µs resolution
1s to 65.535s range: 2ms resolution
Measurement Set-up
Setting
Fill-in-the-blanks using front-panel or
full-size external keyboard
Load settings from floppy disk or via
the LAN port
Program using internal Instrument
BASIC or via GPIB
HELP Function
Library: Default measure setup, Vce-Ic,
Vds-Id, Vgs-Id, and Vf-If are pre-
defined softkeys
User-defined measurement setup
library
Auto file load function on power-up
Measurement
The 4155C and 4156C can
perform dc or pulsed force/measure,
and stress force. For dc,
voltage/current sweep and sampling
(time domain) measurements are
available.
Voltage/Current Sweep
Measurement Characteristics
Each SMU and VSU can sweep using
VAR1 (primary sweep), VAR2
(subordinate sweep), or VAR1
|
(synchronous sweep).
VAR1
Primary sweep controls the staircase (dc
or pulsed) voltage or current sweep.
Maximum number of steps: 1001 for one
VAR1 sweep.
Sweep type: linear or logarithmic
Sweep direction: Single or double sweep
Hold time:
Initial wait time or wait time after
VAR2 is set: 0 to 655.35s with 10ms
resolution
Delay time:
Wait time from VAR1 step to the start
of the measurement: 0 to 65.535s with
100 µs resolution
Functions
VAR1
Staircase or pulse sweep synchronized
with the VAR1 sweep. Sweep is made
with a user specified ratio and offset
value. VAR1
|
output is calculated as
VAR1
= a × VAR1 + b, where a is the
user specified ratio and b is the user
specified offset value.
Linear scale (no limit mode), log
scale, and thinned-out modes:
560µs (720µs at thinned-out mode)
to 1s range:
80µs resolution
1s to 65.535s range:
2ms resolution
Note: The following conditions must be set
when initial interval is less than
2ms.
N
umber of measurement channels: 1
Measurement ranging: fixed range
Stop conditi
on
: disable
Hold time:
Initial wait rime: 0.03s to 655.35s,
100µs resolution
Sampling measurement stop condition:
A condition to stop the sampling can
be defined.
Sampling interval setting accuracy
(supplemental data):
0.5% + 10µs (sampling interval
480µs)
0.5% + 10µs (480µs sampling
interval <2ms)
0.5% + 100µs (2ms sampling
interval)
C-V Measurement
Characteristics
Capacitance is calculated value
derived from the following
equation:
dQ
C =
dQ is the charge when dV, the step
voltage, is applied by the SMU;
dQ is derived from the measure-
ment current (amps) and the
integration time (seconds).
Measurement Range:
Measurement range depends on the
current range, integration time, and
step voltage (refer to the chart in
supplemental data).
Capacitance Calculation Accuracy:
Accuracy is dependent on accuracy of the
current measurement and voltage measure-
ment, and the stray capacitance and
leakage current of measurement system etc
(Refer to the chart in supplemental data).
Zero Offset:
Cancels stray capacitance of the fixtures
and test leads.
Leakage Current Compensation:
Cancels the influence of the leakage cur-
rent to the capacitance measurement.
dV
SMU, VSU, and PGU output can be
forced for the user specified period.
Stress time set range:
5000µs to 31,536,000s (365 days)
Resolution:
100µs (500µsstress time10s)
10ms (10s<stress time31,536,000s)
Burst pulse count:
1 - 65,535 (PGU only)
Trigger:
The 4155C and 4156C output a gate
trigger while stress channels are
forcing stress.
Stress Force Characteristics
|
|