14 Utilities
396 InfiniiVision Oscilloscopes User’s Guide
3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on the rear
panel to the BNC barrel connector.
4 Press the [Utility] key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel CALIBRATION switch to
PROTECTED.
User Cal Status
Pressing [Utility]&Service&User Cal Status displays the summary results of the
previous User Cal, and the status of probe calibrations for probes that can be calibrated.
Note that passive probes do not need to be calibrated, but InfiniiMax probes can be
calibrated. For more information about calibrating probes see page 78.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status: