Seitenansicht 241
Normalization to 0 Hz Bandwidth
When you enter a Normalized Bandwidth of 0 Hz, the system plots the
measured noise data without normalizing it. Plotting the measured data
without normalization, causes all spurs, marked or unmarked to be plotted
at their true level in dBc. The criteria for marking spurs is a function of
the number of averages and the level of the spur. Spurs that were not
marked in a 1 Hz Normalized Bandwidth can be observed at their actual
level when there's no normalization of the noise.
When the measurement data is not normalized (0 Hz bandwidth), it is
plotted in the actual bandwidth in which it was measured. As a result,
breaks in the noise curve will appear at the band edges as shown in
Figure 4-1.
HP304BR DEMO: 10 MHz "FT -s. LQ MHz
M
B'
[hp] :J04BR Carrier: 10.E46 Hz 33 Oc-t 199? 15H5:54 - 15:40:42
1 1 r
-10
-30 -
-3 LI
-40 U.
-60 -
-70 -
-50 P
-90
-L00
-I 10
-L20
-130
-149
-L50 -
-160 h
-170
^
?v^W
i i i i t i ill
100
L K
10K
10
£Cf) Ed Be] vs f[HzH 1B8K
Figure 4-1. Noise Data Plotted in a 0 Hz Bandwidth.
Normalized Bandwidth 4-17