Agilent Technologies HP 70910A Betriebsanweisung Seite 34

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 154
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 33
34 Chapter 2
Installation
Preparing the 11974 for Use
Installation
Figure 2-1 Voltage Selection Switch and Line Fuse Locations
Checking the Fuse
The type of ac-line input fuse depends on the input-line voltage. Use the following fuses:
The line fuse is housed in a small container located inside the rear-panel power-connector
module of the 11974-60028 Preselector Power Supply. When replacing a fuse, be sure that
the arrows on the container match the direction of the arrows on the power-connector door.
Refer to Figure 2-1 on page 34.
100 V to 120 V Operation: 1 A 250 V UL/CSA (TIME DELAY) (part number 2110-0674) (for
use in the United States)
220 V to 240 V Operation: 0.5 A 250 V IEC (TIME DELAY) (part number 2110-0458)
(for use in other countries)