Performance Tests
5. 3 dB Bandwidth
Performance Tests
7. Enter the appropriate millimeter-wave source module frequency multiplication factor
into the synthesized sweeper by pressing
SHIFT and START FREQ. Press the following
keys (on the synthesized sweeper) for the appropriate band:
for an 11974A:
2 Hz
for an 11974Q: 3 Hz
for an 11974U: 3 Hz
for an 11974V: 4 Hz
8. Set the synthesized sweeper for a CW signal at the frequency listed below:
for an 11974A: 26.5 GHz
for an 11974Q: 33.0 GHz
for an 11974U: 40.0 GHz
for an 11974V: 50.0 GHz
9. Activate external leveling on the synthesized sweeper by pressing
XTAL.
10.Adjust the output power of the synthesized sweeper for a reading of approximately
0 dBm on the 8349B display.
11.Set the spectrum analyzer to the appropriate millimeter band by using the following
key sequence for the appropriate band:
for an 11974A:
SHIFT
for an 11974Q: SHIFT
for an 11974U: SHIFT
for an 11974V: SHIFT
12.Set the spectrum analyzer controls as follows:
SHIFT, 5, MHz, 0, and dB (sets external mixer conversion loss to 0 dB)
FREQUENCY SPAN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 MHz
ENTER dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 dB
13.Set the reference level of the spectrum analyzer by pressing
REFERENCE LEVEL and
entering the corresponding value for the band listed below:
for an 11974A:
40 dBm
for an 11974Q:
40 dBm
for an 11974U:
40 dBm
for an 11974V:
55 dBm