7 Press the Measure Delay softkey. Delay is measured and displayed on
the screen.
Negative delay values indicate the defined edge on channel 1 occurred after
the defined edge on channel 2.
Automatic Delay Measurement Hints
If an edge is selected that is not displayed on the screen, delay will not be
measured.
User thresholds have no effect on automatic delay measurements. Delay is
always measured at the 50% transition point (measurement threshold level).
Automatic Delay Measurement
Figure 2 7
Operating the Measurement/Storage Module
To make delay measurements automatically