Probing High-speed Signals
with the Agilent 86100 Series
of Wide-bandwidth
Sampling Oscilloscopes
Product Note 86100-6
The high bandwidth and low noise of equivalent-time
sampling oscilloscopes provide precise displays of
high-speed signals. Historically, if the signal could
not be routed to the oscilloscope through a high-quality
cable, thus requiring the use of a probe, accurate
measurements became difficult to achieve. Three
important measurement accessories help make probe
based measurements both simple and accurate for
the Agilent 86100 Wide Bandwidth Oscilloscope:
Agilent 113X, 115X, and 116X series of high
performance probes with up to 13 GHz of bandwidth
Agilent N1022A probe adapter
Agilent 83496A/B clock recovery module
Seitenansicht 0
1 2 3 4 5 6 7

Inhaltsverzeichnis

Seite 1 - Sampling Oscilloscopes

Probing High-speed Signalswith the Agilent 86100 Series of Wide-bandwidth Sampling OscilloscopesProduct Note 86100-6The high bandwidth and low noise o

Seite 2 - Probe amplifier

ProbesAgilent offers a wide variety of probes with the bandwidthand connectivity required for high-speed waveform analysis. These include differential

Seite 3

The N1022A is directly compatible with the following modules:86101A 86102A and U86103A and B 86105A 86106A 86112A83481A 83483A83485A 83486A83487A 5475

Seite 4

Triggering the oscilloscope with a clockextracted from the probed signalSampling oscilloscopes are different from real-time oscilloscopes in that they

Seite 5 - Probe calibration procedures

5861XX83496AData inData inData outData outProbeN1022AFigure 4: Connector scheme using the 83496A/B and N1022AProbe calibration proceduresThe 86100 DCA

Seite 6 - 83496A/B (Option 101)

6Probing configuration when using theoptical/electrical configuration of the83496A/B (Option 101)The primary usage of the 83496A/B Option 101 is for e

Seite 7 - Remove all doubt

www.agilent.comOnline Assistancewww.agilent.com/find/dcaFor more information on Agilent Technologies’ products, applications or services, please conta

Kommentare zu diesen Handbüchern

Keine Kommentare