Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
Check the setup/hold with single clock, multiple clock
edges
1 Select the logic analyzer setup/hold time.
a In the Analyzer setup window select the Sampling tab.
b Under the Sampling tab, select and activate a rising and falling edge for
any clock.
The Setup/Hold window requires a double clock edge before it will allow a setup/hold
of 5.0/-2.0 ns.
c Select the Format tab. Under the Format tab, select Setup/Hold.
d In the Setup and Hold window, ensure All bits is selected.
e Enter the setup time of the setup/hold combination to be tested in the
Setup: field.
Setup/Hold Combinations
QV
QV
f Select the close (X) button in the upper-right corner to close the Setup/
Hold window.