Chapter 3: Testing Performance
To Test the Time Interval Accuracy
g In the Pod Threshold window, select the Standard threshold voltage
field. At the pop-up menu select ECL (–1.30 V).
3 Set up the Waveform window.
a In the Analyzer setup window, select Window, then select Slot n:
Analyzer<n> (where “n” is the slot you have the module installed), then
select Waveform. A Waveform window opens.
b In the Waveform window select the Markers tab.
c Select the G1 field and a Marker Setup window appears.
d Ensure that the Interval Time field reads “from G1 to G2” (instead of
“from G2 to G1”).
Leave this window open as you will be using it later when acquiring data.