Connecting Instruments to LAN 5
Agilent Connectivity Guide 125
.Using the route add Command If you are having difficulty with a
private LAN (using a hub, switch, or even a simple crossover cable), a
possible cause for no communication between the PC and connected
instruments is that the PC has not recognized that it is connected to two
different subnets. (See Local and Remote Networks for details.)
One way to put the PC and the instrument on the same subnet is to use
route add <Instrument IP Address> <PC IP Address> from the Command
Prompt
. If this action does not correct the problem, go to Web Browser
Checks.)
Example: Using the route add Command
For example, to send a
route add command, open a Command Prompt
window. When the
Command Prompt window opens, type route add
<Instrument IP Address> <PC IP Address>
, where <Instrument IP_address>
is the IP address of the instrument and
<PC IP Address> is the IP address
of the instrument. Then press the
Enter key.
For example, if the IP address of your PC is 155.139.103.152 and the
instrument address is 169.254.58.10, the two devices are probably on
different subnets. To add the instrument to the subnet for the PC, use
one of the following:
! route is lost when the PC is rebooted
route add 169.254.58.10 155.139.103.152
or
! route persists when the PC is rebooted
route -p add 169.254.58.10 155.139.103.152
NOTE
If your devices are connected to site LAN, do not use the route add
command.