Agilent Technologies DisplayPort Wartungshandbuch Seite 139

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 152
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 138
Calibrating the Infiniium Oscilloscope and Probe 14
U7232A DisplayPort Compliance Testing Notes 139
c Follow the on- screen instructions.
d At the end of the Atten/Offset Calibration perform the Skew
Calibration.
Differential Probe Head Skew Calibration
This procedure ensures that the timing skew errors between channels are
minimized. Perform the following steps:
1 Referring to Figure 57 below, perform the following steps:
a Select the Start Skew Calibration button and follow the on- screen
instructions. For more information on proper connection of probe to
the oscilloscope, refer to the De- skew and Calibration manual. This
Figure 56 Probe Calibration Window.
Seitenansicht 138
1 2 ... 134 135 136 137 138 139 140 141 142 143 144 ... 151 152

Kommentare zu diesen Handbüchern

Keine Kommentare