Seitenansicht 40
Module 10
Creating Your Test Definitions
10-21
To Define Test Output
Analysis Parameters
Define Analysis Parameters…
Define the analysis parameters as shown below:
At the Vector Parameters area or the Scalar Parameters area:
1. Click the Add button.
2. Enter the Name, Unit, and Description of the parameter.
For the sweep output/measurement data, use the vector parameter.
For the spot output/measurement data, use the scalar parameter.
For the data given by the max( ) and min( ) built-in function, use the vector parameter.
In this example, VG, ID, and GM are the vector parameter and VTH is the scalar parameter.
GM_MAX is the vector parameter because the max( ) built-in function is used to calculate this value.