Chapter 3: Testing Performance
Performance Test Record
41
DC Gain Accuracy Test
Vertical
Sensitivity
V
Cal Out
Setting
V
DMM+
V
DMM-
V
Scope+
V
Scope-
Calc. DC
Gain Error
DC Gain Error
Test Limits
Channel 1
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 2
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 3
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 4
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 1R
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 3R
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1 V/div ±2.4 V ±2 %