Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
30
Connections for all oscilloscope models, channels 1- 4
Notes
• Connect output 1 of the 11667C splitter to the scope Channel n input directly using the 11901A adapter
and a connector saver, without any additional cabling or adapters.
• Connect the power sensor directly to output 2 of the power splitter using the 11901C adapter without
any additional cabling or adapters.
• Connect the microwave cable directly to output 3 of the power splitter using the 11901C adapter
without any additional cabling or adapters.
• Minimize the use of other adapters.
• Ensure that 2.4 mm and 3.5 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 3.5 mm
8 in-lbs (90 N-cm) for 2.4 mm
Power sensor
cable
Power sensor
N1922A
Power splitter
11667C
11901C adapter
11901A adapter
Microwave cable
Microwave CW
Generator E8257D
Power meter
N1912A
Connection saver
Oscilloscope