Agilent N516xA, N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 189
Programming the Status Register System
Status Groups
Data Questionable Modulation Event Enable Register
The Data Questionable Modulation Event Enable Register lets you choose which bits in the Data
Questionable Modulation Event Register set the summary bit (bit 7 of the Data Questionable
Condition Register) to 1.
Command:
STATus:QUEStionable:MODulation:ENABle <value> where <value> is the sum of the decimal values
of the bits you want to enable.
Example: Enable bit 4 and bit 3 so that whenever either of those bits are set to 1, the Data Questionable
Modulation summary bit of the Data Questionable Condition Register is set to 1. Send the command
STAT:QUES:MOD:ENAB 520 (16 + 8).
Query: STATus:QUEStionable:MODulation:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:MODulation:ENABle <value> command.