1. To optimize phase noise use 5 dBm ± 2 dB.
2. 19 ms (typ) when stepping from greater than 3.2 GHz to less than 3.2 GHz.
External reference
Frequency 10 MHz only
Lock range ±1.0 ppm
Reference output
Frequency 10 MHz
Amplitude > +4 dBm into 50 Ω load (typ)
External reference input
Amplitude 5 dBm ± 5 dB
1
Input impedance 50 Ω (nom)
Step (digital) sweep
Operating modes
Step sweep of frequency or amplitude or both (start to stop)
List sweep of frequency or amplitude or both (arbitrary list)
Sweep range
Frequency sweep Within instrument frequency range
Amplitude sweep Within attenuator hold range (see “Output” section)
Dwell time 1 ms to 60 s
Number of points
Step sweep 2 to 65535
List sweep 2 to 1601 per table
Triggering Auto, external, single, or GPIB
Settling time Standard Opt UNX Opt UNY
Frequency
2
< 9 ms (typ) < 9 ms (typ) < 24 ms (typ)
Amplitude < 5 ms (typ) < 5 ms (typ) < 5 ms (typ)
4