Agilent Technologies 16700 SERIES LOGIC ANALYSIS SYSTEM 16700 Installationsanleitung Seite 98

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 146
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 97
98
Trigger the Oscilloscope with the Timing
Analyzer
Triggering the oscilloscope with the timing analyzer is an “intermodule
measurement,” which means it involves more than one instrument. In
this exercise, the timing analyzer is used to trigger on the symptom,
and the oscilloscope is used to monitor the cause. The symptom we will
trigger on is the glitch. This capturing of data simultaneously is done by
telling the oscilloscope to trigger after the timing analyzer triggers. In
other words, the timing analyzer will “arm” the oscilloscope.
Seitenansicht 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 145 146

Kommentare zu diesen Handbüchern

Keine Kommentare