Agilent Technologies 16700 SERIES LOGIC ANALYSIS SYSTEM 16700 Installationsanleitung Seite 58

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 146
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 57
58
Introduction to State Analysis: Trigger on an
Event
If you’ve never used a state analyzer, you may think it’s a complex
instrument that would take much time to master. Actually, the state
analyzer is not any more difficult to understand than a timing analyzer.
The major difference between a timing analyzer and a state analyzer is
the source of the sample clock. While the timing analyzer samples at
regular intervals using the internal clock, the state analyzer samples
when you tell it to using an external clock. Each time the state analyzer
receives a state clock pulse, the analyzer samples and stores the logic
state of the system under test.
DATA
1. AA
2. 01
3. 2F
4. 3B
AA 01 2F 3B
DATA
CLOCK
Sample Sample Sample Sample
Seitenansicht 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 145 146

Kommentare zu diesen Handbüchern

Keine Kommentare