To remove the display
1 Remove the front panel.
2 Disconnect the ribbon cable and the probe compensation cable from
the display.
3 Using the T10 TORX driver, remove the two screws that hold the
display to the deck.
Make sure that when you reinstall these screws that you use the
correct parts. If longer screws are used, they can short the system
board to ground.
4 As you lift the display, rotate it off the two tabs on the side of the
deck.
To remove the system board
1 Using the T10 TORX driver, remove the six screws that hold the
system board to the deck.
When you reinstall these screws, the long screw and spacer are used between
the two large heatsinks.
2
Remove the three probe sense nuts from the front panel BNC’s.
3 Remove the two screws from the rear-panel interface connector and
the nut from the rear-panel BNC.
4 Disconnect the three ribbon cables and the probe compensation
cable.
5 As you remove the system board, rotate the system board so that the
BNCs clear the front panel.