To perform the self-calibration
In this procedure you load the default calibration factors to give a known
starting point for the firmware calibration. However, once the default
calibration factors are loaded, you must perform the remainder of the
firmware calibration to maintain the accuracy of the oscilloscope.
Table 4-10
Equipment Required
Equipment Critical specifications Recommended
Model/Part
Cable BNC, 3 feet Agilent 10503A
Cable BNC, 9 inches, Qty 2 Agilent 10502A
Adapter BNC tee (m) (f) (f) Agilent 1250-0781
Adapter BNC (f-f) Agilent 1250-0080
1 Check the rear panel CALIBRATOR output level.
If you are not sure how to check the CALIBRATOR, see "To check the output
of the
CALIBRATOR," on page 4-6.
2
Load the default calibration factors.
a Set the rear-panel CALIBRATION switch to UNPROTECTED
(up position).
b Press Print/Utility , then press the Service Menu softkey, then
press the Self Cal Menu softkey.
c Press the Load Defaults softkey.
Self-calibration hint
The instrument is self-calibrated at the factory. However, it should be
self-calibrated again in its working environment after a 30-minute warmup to
obtain the best accuracy.
Service
Adjusting the Oscilloscope
4–25