14
Calibration factor and reflection
coefficient
Calibration factor (CF) and reflection coefficient
(Rho) data is unique to each sensor. The CF
corrects for the frequency response of the
sensor. The reflection coefficient (Rho or )
relates to the SWR based on the following
formula:
Maximum uncertainties of the CF data are
listed in the following table. There is only one
set of CF data used for both high and low
power paths of each sensor.
The uncertainty analysis for the calibration
data was done in accordance with the ISO
Guide. The uncertainty data reported on the
calibration certificate is the expanded
uncertainty with a 95% confidence level and
a coverage factor of 2.
Frequency
U2000A sensor
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
U2001A sensor
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 6 GHz
U2002A sensor
50 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 24 GHz
U2004A sensor
9 kHz to 2 GHz
2 GHz to 6 GHz
U2000B sensor
10 MHz to 2 GHz
2 GHz to 12.4 GHz
12.4 GHz to 18 GHz
U2001B sensor
10 MHz to 2 GHz
2 GHz to 6 GHz
U2000H sensor
10 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
U2001H sensor
10 MHz to 6 GHz
U2002H sensor
50 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
18 GHz to 24 GHz
Uncertainty (%)
(25 °C ± 10 °C)
1.8
1.6
2.0
2.2
2.2
1.8
1.6
2.0
2.0
2.5
2.7
2.7
3.0
1.8
1.8
1.8
2.0
2.2
1.8
2.0
2.0
2.0
2.2
2.0
2.5
2.5
2.7
3.0
Specifications (continued)
SWR =