Agilent U2000 Series USB Power SensorsData SheetThe compact, low-cost alternative to conventional power measurement solutions
10Specifications (continued)1. This accuracy is essentially a combination of linearity, instrumentation accuracy, and traceability to absolute accura
11Specifications (continued)10.750.50.250–0.25–0.5–0.75–1Typical power accuracy at 25 °C for U2000/1/2/4A sensors1–40 –30
12Zero and measurement noisePower range 1U2000/1/2/4A sensors–60 dBm to –35 dBm–38 dBm to –15 dBm–20 dBm to –9 dBm–11 dBm to –5 dBm–7 dBm to 15 dBm10
13Settling timeManual filter, 10-dB decreasing power step (not across switching points)10.0450.04220.090.0540.170.0980.340.17160.660.34321.30.66642.61
14Calibration factor and reflection coefficientCalibration factor (CF) and reflection coefficient (Rho) data is unique to each sensor. The CF corrects
15Specifications (continued)Gate1ParameterSampling rateSweep and OffsetResolutionPowerMeasurement speed 2Performance1.47 Msps0 s < Sweep time <
16General specificationsPhysical characteristicsDimensions (LxWxH)WeightOperating environmentTemperature HumidityAltitudePollutionStorage and shipment
17Display units:Absolute: Watts or dBmRelative: Percent or dBDisplay resolution:Resolution of 1.0, 0.1, 0.01 and 0.001 dB in log mode; one to four dig
181. Available in Agilent Automation-Ready CD2. Agilent IO Libraries Suite 15.0 is required if PC is running on Microsoft Windows Vista 32-bit edition
19Ordering informationBNC (m) to SMB (f) trigger cable, 1.5 m, 50 USB 2.0 Type A to 5-pin Mini-B cable with secure locking mechanism, selectable lengt
“Agilent’s power meters have long been recognized as the industry standard for RF and microwave power measurements.”Why Agilent’s power meters and sen
20Ordering information (continued)1. Refer to www.agilent.com/find/e5813a for more information on the device.Accessories, calibration and documentati
Remove all doubtOur repair and calibration services will get your equipment back to you, performing like new, when prom-ised. You will get full value
3Compact solutions for testing today's RF and microwave communication systemsFor installation and maintenance of base stationsFor production test
4Introducing the U2000 Series USB power sensorsCompact “power meters”, simple set-upThe U2000 Series are standalone sensors. That means they essential
5Introducing the U2000 Series USB power sensors (continued)Faster production testing of multiple channels The U2000 Series’ fast measurement speed of
6Take a closer lookDiode-based sensors frequently rely on the application of correction factors to extend their dynamic range beyond their square-law
7SpecificationsSpecifications contained in this chapter are valid ONLY after proper calibration of the power sensor and apply to continuous wave (CW)
8Typical SWR for U2000A sensor (25 °C ± 10 °C)Specifications (continued)Frequency GHzSWR1.121.101.081.061.041.021.00U2000A024681012141618Typical SWR f
9Specifications (continued)Typical SWR for U2002H sensor (25 °C ± 10 °C)Typical SWR for U2000B sensor (25 °C ± 10 °C)Typical SWR for U2001B sensor (25