Performance and Specifications 7
N9360A Multi UE Tester W-CDMA User Manual 7-3
Frequency Error
Table 7-2 shows the specification of the frequency error
measurement.
Error Vector Magnitude (EVM)
Table 7-3 shows the specification of the EVM measurement.
Table 7-2 Frequency Error
Item Specification Unit Remarks
Measurement Range 0 to ±500 Hz
Measurement
Resolution
0.1 Hz
Display Resolution 0.1 Hz
Accuracy ± (10 +
Reference signal
accuracy)
Hz Downlink Output Level:
–50 dBm
Average of two
measurements.
Input Level Range –20 to +36 dBm
Table 7-3 Error Vector Magnitude (EVM)
Item Specification Unit Remarks
Measurement Range 0 to 20 %
Measurement Resolution 0.01 %
Display Resolution 0.01 %
Residual EVM 3.8 % Downlink Output Level: –50dBm
Input Range –20 to +36 dBm