Agilent PXT Wireless Communications Test Set
User’s Guide
115
onDurationTimer
Enables you change the setting of the on duration timer that is present in the RRC Connection
Reconfiguration message of the currently loaded scenario file.
Overwrites Scenario File Value
Range psf1 | psf2 | psf3 | psf4 | psf5 | psf6 | psf8 | psf10 | psf20 | psf30
| psf40 | psf50 | psf60 | psf80 | psf100 | psf200
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More
drx-InactivityTime
Specifies the number of PDCCH-subframe(s) after successfully decoding a PDCCH that the UE will listen to
the PDCCH before turning off its receiver.
Specifies the number of consecutive PDCCH-subframe(s) after successfully decoding a PDCCH indicating
an initial UL or DL user data transmission for this UE.
For example, a value of psf5 corresponds to 5 PDCCH sub-frames.
Overwrites Scenario File Value
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More
drx-retransmission timer
Overwrites Scenario File Value
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More