
Chapter 3 73
Programming Command Reference
TESTs Subsystem for AUTOMATIC TEST
3. Programming Command
Reference
TESTs:AUTO:RESults:RAMP6:ZOOM ON|OFF|1|0
TESTs:AUTO:RESults:RAMP6:ZOOM?
This command activates the zooming function of the Power Ramp graph for each
of the six measurements. The choices are ON (1) and OFF (0). If set to ON, the
rising and trailing edges are expanded. At preset, these are set to OFF.
Power Ramp Marker Measurement
TESTs:AUTO:RESults:RAMP1:MARKer?
TESTs:AUTO:RESults:RAMP2:MARKer?
TESTs:AUTO:RESults:RAMP3:MARKer?
TESTs:AUTO:RESults:RAMP4:MARKer?
TESTs:AUTO:RESults:RAMP5:MARKer?
TESTs:AUTO:RESults:RAMP6:MARKer?
This query command returns a set of marker position values, <string 1>,
<string 2>, <string 3> and <string 4>, in the Power Ramp graph for
each of the six measurements.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• <string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result.
• <string 3> (7 digits) is the marker position ranging from −009.25 to +156.25
(bits).
• <string 4> (5 digits) is the power level of the marker position ranging from
−99.9 to +99.9 (dB). If any measurement is overranged, underranged or not
made, the value −99.9 is returned.
Power Ramp Marker Position Control
TESTs:AUTO:RESults:RAMP1:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:RAMP1:MARKer:POINt <real>
TESTs:AUTO:RESults:RAMP1:MARKer:POINt?
TESTs:AUTO:RESults:RAMP2:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:RAMP2:MARKer:POINt <real>
TESTs:AUTO:RESults:RAMP2:MARKer:POINt?
TESTs:AUTO:RESults:RAMP3:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:RAMP3:MARKer:POINt <real>
TESTs:AUTO:RESults:RAMP3:MARKer:POINt?
TESTs:AUTO:RESults:RAMP4:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:RAMP4:MARKer:POINt <real>
TESTs:AUTO:RESults:RAMP4:MARKer:POINt?
TESTs:AUTO:RESults:RAMP5:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:RAMP5:MARKer:POINt <real>
TESTs:AUTO:RESults:RAMP5:MARKer:POINt?
Kommentare zu diesen Handbüchern