20
Completing the ROM upgrade
Upon initial power
up
When new ROMs are installed in the HP/Agilent 8922, and the
instrument is first powered on, there may be a delay of up to two minutes
before anything appears on the display. Do not power cycle during this
time.
When the instrument successfully powers on, the following three
messages appear:
• Cal file checksum incorrect. File reset to default
values.
• All Self Tests Passed.
• Frequency Reference Cal Lost. Perform Reference
Calibration.
The first two messages appear and disappear quickly as they are
replaced by the third message. To view all messages, use the message
(MSSG) screen.
To clear RAM It is good practice to clear the instrument RAM following the instrument
upgrade. To do this:
Step 1. Move the cursor to the CONFIG field and select.
Step 2. In the CONFIGURE screen, move the cursor to the SERVICE field and
select.
Step 3. In the SERVICE screen, locate the RAM Initialize field and select.
Step 4. Select on the front panel keypad to clear the registers.
YES