Agilent Technologies 8753ET Betriebsanweisung Seite 208

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 546
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 207
1-194 Chapter 1
Alphabetical Command Reference
RAWOFFS
RAWOFFS
Syntax
RAWOFFS<ON|OFF>; or RAWOFFS?;
Description
By turning raw offsets off (RAWOFFSOFF), a full two-port error correction can be performed
without including the effects of the offsets. It also saves substantial time at recalls and
during frequency changes (see SM8). Raw offsets follow the channel coupling. RAWOFFS and
SAMC are linked in that when one changes state, so does the other. See “Example 2G: Take4
— Error Correction Processed on an External PC” on page 7-48.
Front Panel Equivalents
Command Description Range Query Response
RAWOFFS Selects whether sampler and attenuator
offsets are on or off.
N/A
<0|1><
L
F
>
Command Hardkey Softkey
RAWOFFS
System
RAW OFFSET ON OFF
Seitenansicht 207
1 2 ... 203 204 205 206 207 208 209 210 211 212 213 ... 545 546

Kommentare zu diesen Handbüchern

Keine Kommentare