122 Agilent 34980A Mainframe User’s Guide
3 Features and Functions
AC Voltage, AC Current (for all ranges):
Frequency, Period:
Digital Input, Totalize:
Front Panel Operation:
Channel (Configure) > CHANNEL DELAY > AUTO
Once you have added the specified channel to the scan list, the channel
delay choice will be visible in the menu.
Interface Operation: The following command enables an automatic channel
delay on the specified channels.
ROUT:CHAN:DELAY:AUTO ON,(@1003,1013)
Selecting a specific channel delay using the ROUTe:CHANnel:DELay
command (see “Channel Delay” on page 120) disables the automatic
channel delay.
A C F i l t e r C h a n n e l D e l a y
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
7.0 seconds
1.0 second
120 ms
A C F i l t e r C h a n n e l D e l a y
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
600 ms
300 ms
100 ms
Channel Delay
0 seconds
Seitenansicht 129