Agilent Technologies U1270 Series Bedienungsanleitung Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 24
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
7
General Specifications
Display
Liquid crystal display (LCD) (with maximum reading of 33000 counts)
Power consumption
460 mVA maximum (with backlight enabled)
Battery Type
4 × 1.5 V Alkaline battery (ANSI/NEDA 24A or IEC LR03), or
4 × 1.5 V Zinc Chloride battery (ANSI/NEDA 24D or IEC R03)
Battery Life
300 hours typical (based on new Alkaline batteries for dc voltage measurement)
Low battery indicator will flash when the battery voltage drops below 4.4 V (approximately)
Fuse
10 × 35 mm 440 mA/1000 V 30 kA fast-acting fuse
10 × 38 mm 11 A/1000 V 30 kA fast-acting fuse
Operating Environment
Operating temperature from –20 to 55 °C, 0 to 80% RH
Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50%
RH at 55 °C
Altitude up to 2000 meters
Pollution degree II
Storage Compliance
–40 to 70 °C, 0 to 80% RH
Safety Compliance
CAN/CSA-C22.2 No. 61010-1-04
• EN/IEC 61010-1:2001
• ANSI/UL 61010-1:2004
Measurement Category
CAT III 1000 V/ CAT IV 600 V
Electromagnetic Compatibility
(EMC)
Commercial limits compliance with EN61326-1
Ingress Protection Rating
IP-54
Temperature Coefficient
0.05 × (specifi ed accuracy) / °C (from –20 to 18 °C, or 28 to 55 °C)
Common Mode Rejection Ratio
(CMRR)
>120 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
Normal Mode Rejection Ration
(NMRR)
>60 dB at 50/60 Hz ± 0.1%
Dimensions (W x H x D)
92 × 207 × 59 mm
Weight
U1271A: 518 grams (with batteries)
U1272A: 520 grams (with batteries)
Warranty
Three years for product
Three months for product’s accessories
Calibration Cycle
One year
Specification Assumptions
Accuracy is given as ±(% of reading + counts of least signifi cant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH.
AC V and AC μA/mA/A specifi cations are ac coupled, true RMS and are valid from 5% of range to 100% of range.
The crest factor may be up to 3.0 at full- scale except for the 1000 V range where it is 1.5 at full scale.
For non- sinusoidal waveforms, add (2% reading + 2% full scale) typical, for crest factors up to 3.
• After Z
LOW
voltage measurements, wait at least 20 minutes for thermal impact to cool before proceeding with any other measurement.