AC + DC specifications for U1272A
Function Range Resolution
Accuracy
45 Hz to
65 Hz
20 Hz to
2 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
20 kHz
to 100
kHz
True RMS
AC + DC
Voltage
[1]
30 mV 0.001 mV 0.7 + 40 0.8 + 45 1.1 + 45 1.1 + 60 3.6 + 60
300 mV 0.01 mV 0.7 + 25 0.8 + 30 1.1 + 30 1.1 + 45 3.6 + 45
3 V 0.0001 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 3.6 + 45
30 V 0.001 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 3.6 + 45
300 V 0.01 V 0.7 + 25 1.1 + 30 1.6 + 30 2.1 + 45 -
1000 V 0.1 V 0.7 + 25 1.1 + 30 1.6 + 30 - -
Function Range Resolution
Accuracy
Burden
voltage/
Shunt
45 Hz to
65 Hz
20 Hz to
2 kHz
True RMS
AC + DC
Current
[2]
300 µA 0.01 µA 0.8 + 30 1.1 + 30 <0.04 V/100 Ω
3000 µA 0.1 µA 0.8 + 30 1.1 + 30 <0.4 V/100 Ω
30 mA 0.001 mA 0.8 + 30 1.1 + 30 <0.08 V/1 Ω
300 mA 0.01 mA 0.8 + 30 1.1 + 30 <1.00 V/1 Ω
3A 0.0001 A 0.9 + 35 1.3 + 35 <0.1 V/0.01 Ω
10 A 0.001 A 0.9 + 35 1.3 + 35 <0.3 V/0.01 Ω
1. Notes for voltage specifications:
Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10%
of range.
2. Notes for current specifications:
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse
Specification for 300 mA range: 440 mA continuous.
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.