Figure 18 WiMAX waveform duration 5 ms for UL in TDD mode
Figure 19 Frame duration (5 ms) of UL waveform captured from Signal Studio
Measuring Uplink (UL) Wave-
form
The Signal Studio under Uplink step
reveals that the frame duration of 5 ms
(see Figure 19). The frame durations
are exactly same as the DL waveform
setup except for the Uplink region that
shows the –0.15 dB when only UL
Burst 1 is ON.
Instructions Keystrokes
On the P-Series:
1. Select the WiMAX predefi ned
preset value.
[Preset], Press [
or ] key to
choose {WiMAX}. [Select], {Confi rm}.
2. Set the trace display. [Disp], {Disp Type}, {Trace}.
3. Set the trigger holdoff 4 ms. [Trig/Acq], {Settings}, {1 of 2}, {Hold-
off 4 ms}
4. Expand the window. Change the
scale. Move the marker to
measure 5 ms waveform duration
(See Figure 18).
[ ], {Trace Control}, {X Start
1.5 ms, X Scale 600 µs & Y Scale
10 dB}, {Gate Control}. Press [
] to allocate marker 2 and [ ] to
allocate marker 1.