Chapter 1 19
General Information
Required Equipment
1. General Information
Cable BNC(m)-BNC(m) Cable, 30 cm
BNC(m)-BNC(m) Cable, 122 cm
Coaxial cable with Type-N (m)
connectors, 61 cm (24 in), 2 ea.
Coaxial cable with 75 Type-N (m)
connectors, 61 cm (24 in), 2ea
Agilent p/n 8120-1838
Agilent p/n 8120-1840
Agilent N6314A (p/n
8120-8862)
Agilent p/n 8120-8898
1
1
1
1
P
P,A
P,A ,T
P,A ,T
Adapter N(m)-N(m) 50 Adapter
N(m)-N(m) 75 Adapter
N(m) 50 - N(f) 75 Adapter
N(m)-BNC(f) Adapter
BNC(m)-BNC(m)
BNC Tee (m) (f) (f)
Agilent p/n 1250-0778
Agilent p/n 1250-1528
Agilent p/n 1250-0597
Agilent p/n 1250-0780
Agilent p/n 1250-0216
Agilent p/n 1250-0781
1
1
1
1
1
1
A,T
A,T
P
P,A
P
P
Torque Wrench Size: 3/4 inch
Torque: 136 N-cm
Agilent p/n 8710-1766 1 P,A
USB/GPIB Interface No Substitute Agilent 82357A 1 P,A
Personal Computer
with GPIB board
Windows 2000 / XP, VEE 6.2 or
VEE 7.xx
1P,A
*1.P: Performance Tests, A: Adjustment, T: Troubleshooting
*2.Opt.050 and Opt.124 can be substituted for Opt.030. In this case, a N(m)-BNC(f) adapter is necessary.
*3.The accuracy of a standard E4419B of 0.02 dB is adequate for all tests/adjustments except “Dynamic Accuracy
Test”. This test requires a power meter with Option G12 or H12 that has been certified to a higher accuracy speci-
fication. If an Option G12 or H12 power meter is not available, a test is provide on “Power Meter Accuracy Test”
on page 20 to verify the accuracy of a standard power meter.
Table 1-2 Alternative Test Equipment
Equipment Critical specifications Model Qty.
Use
*1
Frequency Counter Frequency: 50 MHz to 3 GHz
Accuracy: < 2.5 ppm
Agilent 53132A/181A with
Opt.010 and 030
*2
1P,A
Power Meter No Substitute
Agilent E4419B
*3
1P,A
Calibration Kit No Substitute Agilent 85032F 1 T
Short Type-N(m) Agilent 11512A 2 P,T
*1.P: Performance Tests, A: Adjustment, T: Troubleshooting
*2.Opt.050 and Opt.124 can be substituted for Opt.030. In this case, a N(m)-BNC(f) adapter is necessary.
*3.If an accurate measurement of the dynamic accuracy specification is not required, a standard E4419B can be used.
Table 1-1 Recommended Test Equipment
Equipment Critical specifications Recommended Model Qty.
Use
*1