104 Chapter 4
Troubleshooting
Troubleshooting Using Diagnostic Test
Figure 4-13 Setup dialog box
The test result file named “resultDT.txt” is created on the drive D
(D:\Agilent\Service\Log\) of the E5061A/E5062A after “Exit” button
of the Main Menu is pressed. The text file can be read and edited with
a PC.
Diagnostic Test Failure Troubleshooting
Table 4-3 represents the contents of the diagnostic tests and the relationships of failed tests
to probable faulty board assemblies. If the instrument fails the diagnostic test, replace the
faulty board assembly as shown in Table 4-3.
Table 4-3 Diagnostic tests failure troubleshooting information
Test
No.
Test group Probable faulty board assembly
A1 Source A2 Receiver A25 Handler Correc.
Constant
1 A1 Board DC-BUS
Test
### #
2 A2 Board DC-BUS
Test
###
3 A1 Board Internal
Level Monitor Test
### # #
4 A2 Board Internal
Level Monitor Test
## ## #
5 Source Power
Accuracy (step
Attenuator) Test
### # #
6 Source Power
Linearity (power
Sweep) Test
### #
7 Receiver Absolute
Gain Test
## ### #